Catalog #: 20003800
Intended use
Automated chromogenic assay for the quantitative determination of Antithrombin in human citrated plasma on the ACL Futura/ACL Advance and ACL TOP Systems.
Summary and principle
Antithrombin (AT) or Heparin Cofactor I is the major inhibitor of blood coagulation and is essential for effective heparin therapy. By inhibiting the coagulation proteases, especially thrombin, FXa and FIXa, AT prevents uncontrolled coagulation and thrombosis.
Antithrombin deficiency is associated with a high risk of thromboembolic disorders.
Liquid Antithrombin XL can be used to exclude or diagnose hereditary deficiencyin patients with a tendency toward thromboembolism, in pre-operative stages, before prescription of oral contraceptives, DIC6, nephrotic syndromes, liver diseasesand in therapy with heparin or antithrombin concentrates.
The Liquid Antithrombin XL kit is an assay based on a synthetic chromogenic substrate and on FXa inactivation. As a consequence, the method is specific and not influenced by Heparin Cofactor II. Antithrombin levels in patient plasma are measured automatically on IL Coagulation Systems in two stages:
Composition
Chromogenic substrate (Cat. No. 0020003820): 4 x 2.5 mL vials of the chromogenic substrate S-2765, N-α-Z-D-Arg-Gly-Arg-pNA.2HCl (7.5 mg/vial), surfactant and buffer. Opened reagent is stable: 5 weeks at 2-8°C in the original vial or 3 days at 15°C on the ACL Futura®, ACL Advance and ACL TOPTM Systems. Do not freeze.
Factor Xa reagent (Cat. No. 0020003810): 4 x 6 mL vials of a solution containing bovine Factor Xa (60 nkat/vial), heparin, buffer, bovine serum albumin and preservatives. Opened reagent is stable: 5 weeks at 2-8°C in the original vial or 3 days at 15°C on ACL Futura, ACL Advance and ACL TOP Systems. Do not freeze.
Unopened reagents are stable until the expiration date shown on the vial when stored at 2-8°C. For optimal stability remove reagents from the system and store them at 2-8°C in the original vial.
Additional reagents and control plasmas
The following are not supplied with the kit and must be purchased separately.
| Name | Type | Size |
|---|---|---|
| HemosIL Liquid Antithrombin XL Package Insert | application/pdf | 106.28 kB |
| Instrumentation Laboratory Hemostasis Catalog | application/pdf | 908.14 kB |