Catalog #: 5152005
Reagent kit for carrying out Factor VIII inhibitor assays. For Research use only.
COMPOSITION & REAGENT PREPARATION
The Factor VIII Inhibitor Reagent kit for 2-4 determinations contains:
The reagents for Factor VIII assay are not included.
Residual % Factor VIII Activity
A test sample has no inhibitor if the FVIII value of the test sample is the same as the value of the control mixture i.e. the % Factor VIII activity remaining after incubation is 100%.
Definition of Bethesda Unit
A test sample treated by the method above with a residual Factor VIII activity of 50% contains one Bethesda Unit of Factor VIII Inhibitor/mL.
Calculation
Using semi log graph paper, residual % FVIII activity is plotted against Bethesda Units/mL (Linear Axis). The residual % FVIII activity between 25% and 75% can then be read off and the number of units FVIII Inhibitor/mL is multiplied by the dilution factor to obtain the number of Bethesda units per mL in the sample.
| Name | Type | Size |
|---|---|---|
| Factor VIII Inhibitor Flyer | application/pdf | 1.22 MB |
| Factor VIII Inhibitor Kit Material Safety Data Sheet (MSDS) | application/pdf | 29.67 kB |
| Factor VIII Inhibitor Kit Package Insert | application/pdf | 118.86 kB |
| Technoclone Research Use Only Certification for Customers | application/pdf | 30.79 kB |