Intended use

Lyophilized human plasma, prepared by addition of LMW heparin. Calibrated against the 1st International WHO standard (LMW heparin). For construction of calibration curves for quality control of chromogenic heparin assays.

Summary and principle

Heparin is the most frequently used antithrombotic therapeutic drug. The biological activity of this sulfated glycosaminoglycan resides in its ability to accelerate (up to 2000-fold) the inhibitory effect of antithrombin (AT) on the coagulation proteases.

The Low and High Levels of Control Plasma LMW Heparin are prepared by means of a dedicated process and contain Low and High concentration of LMW Heparin.

They are intended for the assessment of precision and accuracy of the assay at Heparin low and high levels.

Reagents -- Composition, Package, Storage and Stability

Control Plasma LMW Heparin, Low: 4 x 1 ml vials (0.3 IU / ml) of a lyophilized powder of human plasma containing low molecular weight (LMW) heparin.

Control Plasma LMW Heparin, High: 4 x 1 ml vials (0.7 IU / ml) of a lyophilized powder of human plasma containing low molecular weight (LMW) heparin.

Unopened controls are stable until the expiration date shown on the vial when stored at 2-8°C. Stability after reconstitution: 24 hours at 20-25°C, or 14 days at 2-8°C in the original vial.

Related Product: LMW Heparin Calibration Plasma Cat # C823500 (823500).

Downloads

Name Type Size
Chromogenix Coamatic Heparin Packet application/pdf 375.95 kB
Chromogenix Heparin Monograph application/pdf 269.58 kB
Chromogenix Heparin Product Range (1999) application/pdf 240.88 kB
Chromogenix LMW Hepain Control Plasma MSDS application/pdf 114.27 kB
Chromogenix LMW Hepain Control Plasma Package Insert application/pdf 82.51 kB